MAGUIRE Enterprises,
Inc.
1-800-548-9686
Considerations When
'Slaving' Monitors and Defibrillators
"Slaving"
The defibrillator/monitor
Some hospital customers may wish
to interface the defibrillator/monitor to other
medical equipment, such as bedside or central
station monitors. Commonly referred to as "slaving",
the patient's ECG signal is cabled from one
device to another, for the purposes of therapy
or remote monitoring. This document describes
typical interconnection schemes, patient risks
that may be incurred, and how to decrease these
risks.
Review: What is Synchronized
Cardioversion?
In patients with certain arrhythmia,
administering a random counter shock carries
the risk of inducing ventricular fibrillation(1).
In such cases, synchronized cardioversion is
usually indicated. A defibrillation shock delivered
during the vulnerable period of the cardiac
cycle, identified in FIGURE
1, may actually cause ventricular
fibrillation.

FIGURE
1: Timing A Synchronized Counter Shock
to the Cardiac Cycle
With the SYNC
mode on the defibrillator/monitor selected,
each QRS complex the patient produces is identified
and marked. The defibrillator only allows delivery
of a shock immediately after a QRS complex is
detected. The Association for the Advancement
of Medical Instrumentation (AAMI) recommends
that defibrillators operating in synchronized
mode deliver the synchronized shock within 60
ms after the peak of the R wave(2).
Since the vulnerable period immediately follows
this60 ms "window", delivery of the
shock must not occur more than
60 ms after the peak of the R wave.
"Slaving"
A Bedside Monitor From The defibrillator/monitor
(crash Cart)
When a defibrillator/monitor is
used in conjunction with a bedside monitor,
the clinical staff may want to view the patient's
ECG signal on both units. A typical connection
would allow the patient to be connected to the
defibrillator/monitor plus
have the ability to quickly interconnect a cable
to the bedside monitor and it's associated central
station. This is the preferred method
for synchronized cardioversion as delay time
is not an issue.
When a patent requiring synchronized
cardioversion is already connected to another
bedside monitor, it may be impractical or undesirable
to place additional monitoring electrodes on
the patient's chest. To monitor with the defibrillator/monitor
during the cardioversion procedure, the clinical
staff may elect to "slave" an ECG
signal from the bedside monitor to the defibrillator/monitor,
as shown in FIGURE
2. The delivery of defibrillation
therapy with devices interconnected in this
fashion can result in specific patient risks
that must be understood and managed by the hospital's
clinical and biomedical staff.
"Slaving"
The Defibrillator/monitor From Another Monitor
For Synchronized Cardioversion
If the defibrillator/monitor is
to be "slaved" from a bedside monitor
as shown in FIGURE
2, it is critical that the ECG
signal obtained from the bedside monitor is
real time; that is, with minimal delay introduced
by the bedside monitor. The hospital's biomedical
engineering staff should perform sync delay
measurements on the system as a whole to ensure
that the 60 ms limit for sync delay is not exceeded.
For equipment designed to easily measure sync
delay, contact MAGUIRE
Enterprises, Inc. or go to our on-line
catalog.

FIGURE
2: Slaving aA LIFEPAK 12 defibrillator/monitor
From A Bedside Monitor For Synchronized Cardioversion
Cable Availability
For assistance in acquiring the
appropriate cables, contact MAGUIRE
Enterprises, Inc. at 1-800-548-9686 or go
to our on-line catalog.
Be prepared to provide the manufacturer and
model number of the bedside monitor and defibrillator
to be used. Patient cable input adapters are
also available allowing the monitor cable to
be plugged into the defibrillator, even if the
original connectors were non compatible.
Patient Risk Current
When interconnecting medical equipment
as shown in FIGURE
2, chassis leakage or risk
currents may be increased. To ensure patient
safety, the hospital biomedical staff should
perform electrical safety testing on the complete
system to ensure that safe current limits are
not exceeded.
Noninvasive Pacing
IMPORTANT: The defibrillator/monitor
should not be slaved from another monitor for
pacing?
The defibrillator/monitor is primarily
used as a demand-mode pacer, which delivers
pacing therapy based on the ECG signal acquired
through it's patient cable. Placing another
monitor in the ECG signal path (as shown in
FIGURE 2)
may result in unexpected distortions of the
pacer-driven ECG waveform, which may, in turn,
affect pacing therapy. When pacing, the defibrillator/monitor
must be connected directly to the patient through
the patient cable.
1
Crockeett, Paula J., RN, Droppert, Beth M. RN,
Higgins, Sandra, MBA, and Richards, R. Kipper,
RN, MN, "Defibrillation: What You Sholuld
Know", ©1996, Physio-Control Corporation